*El PAS que se detalla aquí no se puede aplicar con un sistema o una versión diferente.
Al aplicar el PAS a su sistema de resonancia magnética, asegúrese de consultar el manual que se puede descargar al mismo tiempo que el PAS.
Axial DWI with Exsper
Sequence | SEEPI2D |
Parallel Acquisition | Exsper 2x1 |
Scan time | 0:38 |
Resolution | 1.6x1.6 mm |
FOV | 23 cm |
Matrix size | 144x144 |
Slice thickness | 5 mm |
Slice gap | 1 mm |
Number of slices | 25 |
TR | 4130 ms |
TE | 77 ms |
NAQ | 1 |
b-value | 1000 |
Axial FLAIR with Exsper
Sequence | FSE2D |
Parallel Acquisition | Exsper 2x1 |
Scan time | 1:40 |
Resolution | 1.0x1.0 mm |
FOV | 23 cm |
Matrix size | 224x224 |
Slice thickness | 5 mm |
Slice gap | 1 mm |
Number of slices | 25 |
TR | 10000 ms |
TE | 104 ms |
NAQ | 1 |
Axial TOF MRA with SPEEDER, Fast 3D mode 50%
Sequence | MRA3D |
Parallel Acquisition | SPEEDER 2x1 |
Scan time | 1:15 |
Resolution | 1.3x0.8 mm |
FOV | 20 cm |
Matrix size | 160x256 |
Slice thickness | 1.3 mm |
Number of slabs | 3 |
Number of slices | 116 |
TR | 14 ms |
TE | 6.5 ms |
NAQ | 1 |
Axial T2*w with Exsper
Sequence | FE2D |
Parallel Acquisition | Exsper 2x1 |
Scan time | 0:53 |
Resolution | 1.0x1.0 mm |
FOV | 23 cm |
Matrix size | 240x240 |
Slice thickness | 5 mm |
Slice gap | 1 mm |
Number of slices | 25 |
TR | 380 ms |
TE | 9 ms |
NAQ | 1 |
Axial T2w with Exsper
Sequence | FSE2D |
Parallel Acquisition | Exsper 2x1 |
Scan time | 0:40 |
Resolution | 0.9x0.9 mm |
FOV | 23 cm |
Matrix size | 256x256 |
Slice thickness | 5 mm |
Slice gap | 1 mm |
Number of slices | 25 |
TR | 4400 ms |
TE | 90 ms |
NAQ | 1 |
Axial T1w with Exsper
Sequence | SE2D |
Parallel Acquisition | Exsper 2x1 |
Scan time | 0:56 |
Resolution | 1.0x1.0 mm |
FOV | 23 cm |
Matrix size | 224x224 |
Slice thickness | 5 mm |
Slice gap | 1 mm |
Number of slices | 25 |
TR | 428 ms |
TE | 8 ms |
NAQ | 1 |
Coil
Octave SPEEDER Head 1.5T
© CANON MEDICAL SYSTEMS ARGENTINA S.A.
© CANON MEDICAL SYSTEMS ARGENTINA S.A.